Digital EEG Mapping
Model:
Minimum order:
Wheelbase:
Weight:
Function:
Feasures:
10-20 Standard EEG Lead System
Observe and record EEG simultaneously
Dynamic 3D rotated brain relief mapping
Flash Stimulation
Audio Stimulation
Adopting Large Scale Integrated circuit (LSI) and improved anti-jam capability.
Doubly isolated the power supply and brain electrical signal to prevent creepage and to insure security.
Function of various event description such as open or close eyes, deep breathing and photic stimulation
Arbitrarily set lead modes such as monopole, dual-pole and reference electrode
Choose any lead to enlarge and automatically measure
Auto-replay EEG wave and real-time compressed chart-array map
Arbitrarily choose histogram, graph, and graph and percentage map
Continuously replay spike wave relief map and dynamic 3D rotated relief mapArbitrarily set digital filtering, formfeed speed, gain, transmission bands and time constant settings
Self-contained medical records management system and manifold report patterns
Windows 7/Vista operating system, easy to operate
Optional multi-parameter leads of ECG, EMG, EOG, Resp. etc.

Technical parameter:
Voltage measurement: 25mV to 200mV, error range: ±10%.
Time interval: error range is no more than ±5.
Time constant: 0.03s to 0.1s, error range: ±40%; the error range is no more than±20% when the time constant ≥0.1s.
Amplitude-frequency characteristic: 1Hz to 30Hz, deviation range: +5% to -30%.
Power spectrum amplitude: error range: ±10%.
Power spectrum frequency: deviation range: ±5%.
Noise level: ≤5µVp-p (peak to peak value).
CMRR (Common mode rejection ratio): each lead ≥80dB.
Polarizing voltage endurance: voltage error range is no more than ±5% when added ±300mV direct current polarizing voltage.