Hospital Icu Patient Monitor With Etco2

Model: CONTEC8000

Minimum order:

Wheelbase:

Weight:

Function: The Hospital icu patient monitor with etco2 has abundant functions that can be used for clinical monitoring with adults, pediatrics, and neonates. Users may select different parameter configurations according to different requirements.

ICU is the intensive care unit, which can be divided into comprehensive ICU, specialized ICU, and emergency ICU. It is a professional ward for the centralized management of critically ill patients, equipped with professional medical staff, and various advanced medical monitoring and treatment equipment. The monitoring equipment often used in ICU is as follows:Common ECG monitors, EEG monitors, ECG, bedside ultrasound, fiberoptic bronchoscopy, X-rays, etc. For trauma-based specialist ICUs, there will also be PICCO monitors, which are mainly used to monitor patients' blood pressure, central venous pressure, and the presence of pulmonary edema
Send Inquiry
Product Description

Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP


1) ECG Heart rate(HR)

    ECG waveform

    Arrhythmia and ST-segment analysis

2) RESP Respiration rate(RR)

    Respiration waveform

3) SpO2 Oxygen saturation(SpO2)

    Plethysmogram(PLETH) waveform

    Pulse rate(PR)

    Bar graph

4) NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean Pressure(MEAN)

5) TEMP T1, T2, TD

6) IBP(optional) CH1: SYS, DIA, MAP

    CH2: SYS, DIA, MAP

    IBP waveform

7) CO2 EtCO2

    InsCO2:Inspired Minimum CO2

    AwRR: Air Way Respiration


Specification: 


ECG

Lead mode: 3-lead or 5-lead

Lead selection: I, II, III, aVR, aVL, aVF, V

Waveform: 5-lead, dual-channel

3-lead, single-channel

Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV

Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s

HR:

Measurement and alarm range: 15~350 bpm

Accuracy: ±1 % or ±1 bpm, whichever is greater

Alarm accuracy: ± 2 bpm

Resolution: 1 bpm

ST-segment monitoring:

Measurement and alarm range: -2.0 mV ~ +2.0 mV

Accuracy: -0.8 mV~+0.8 mV ±0.04 mV or ±10%, whichever is greater

Other range: unspecified

Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,

PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC

Pacemaker: yes

 

RESP

Method: R-F(RA-LL) Impedance

Respiration rate:

Measurement and alarm range: 0~150 rpm

Resolution: 1 rpm

Measurement accuracy: ±2 rpm

Alarm accuracy: ±3 rpm

Apnea alarm: 10~40s

Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s

 

NIBP

Method: Oscillometry

Mode: Manual/AUTO/STAT

Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes

Measurement period in STAT mode: 5 minutes

Measurement and alarm range: 10 ~ 270 mmHg

Resolution: 1 mmHg

Cuff pressure accuracy: ±3 mmHg

Measurement accuracy:

Maximal mean deviation: ±5 mmHg

Maximal standard deviation: 8 mmHg

Over-pressure protection:

Adult mode: 297±3 mmHg

Pediatric mode: 240±3 mmHg

Neonatal mode: 147±3 mmHg

 

SpO2

Measurement and alarm range: 0 ~ 100%

Resolution: 1%

Measurement accuracy: 70%~100%: ±2%;

0%~69%: unspecified

 

PR

Measurement and alarm range: 30 ~ 250 bpm

Measurement accuracy: ±2 bpm or ±2%, whichever is greater

 

TEMP

Channel: dual-channel

Measurement and alarm range: 0 ~ 50℃

Resolution: 0.1 ℃

Accuracy: ±0.1 ℃


Accessories: 

1) Adult fingertip SpO2 probe (5-pin)

2) Adult NIBP cuff

3) NIBP extension tube

4) ECG lead cable

5) ECG electrode

6) Temperature probe

7) Power cord

8) User Manual (GB only)on request for other Language

9) Thermal Printer

10) ETCO2