12.1 inch patient monitor, table patient monitor, ambulance patient monitor
Model: ML-PM80C
Minimum order:
Wheelbase:
Weight:
Function:
12.1-inch high-resolution color TET LCD display.
Thinner and lighter in design, convenient and portable for users.
Real-time S-T segment analysis, pacemaker detection.
Various interfaces: standard screen, trend screen, oxy CRG screen, NIBP list screen, big front screen.
3-level audio/visual alarms.
Efficient resistance to interference of defibrillator and HF knife.
Up to 400 groups NIBP list, 6000seconds, ECG waveform recall, 60 alarm event records recall, 7-day trend chart in storage.
Built-in rechargeable lithium battery, 2-hour working capability.
Wireless central monitoring system applicable to ICU/CCU/OR.
Description :
NIBP :
Method: Oscillometry
Measure mode: Manual, Auto, STAT
Measure interval in auto mode : 1~5/10/15/30/60/90/120/180/240/480min
Measure period in STAT mode : 5min
Pulse rate range: 40 ~ 240bpm
Measure & alarm range:
Adult :
SYS 40 ~ 280mmHg
DIA 10 ~ 220mmHg
MEAN 20 ~ 240mmHg
Pediatric :
SYS 40 ~ 220mmHg
DIA 10 ~ 160mmHg
MEAN 20 ~ 170mmHg
Neonatal :
SYS 40 ~ 135mmHg
DIA 10 ~ 100mmHg
MEAN 20 ~ 110mmHg
Static pressure measurement
Range: 0 ~ 300mmHg
Accuracy: ± 3mmHg
Resolution: 1mmHg Accuracy:
Max. Mean Error: ± 5mmHg
Max. Standard Deviation: 8mmHg
Overpressure protection:
Adult 300mmHg
Pediatric 240mmHg
Neonatal 150mmHg
SPO2
Measurement range: 0 ~ 100%
Resolution: 1%
Accuracy: 70% ~ 100% ± 2%
40% ~ 69% ± 3%
Environment :
Operating environment :
Temperature: 0 ~ 40 ℃
Humidity: 15% ~ 80%
Pressure: 86.0 kPa ~ 106.0kPa
Transport and Storage environment :
Temperature: -20 ~ 55 ℃
Humidity: 10% ~ 90%
Pressure: 86.0 kPa ~ 106.0kPa
Send Inquiry